H2S-B4 High Resolution Hydrogen Sulfide Sensor (H2S sensor) Is Used To Monitor Trace Hydrogen Sulfide In Air Quality

Categories Gas Sensor
Brand Name: alphasense
Model Number: H2S-B4
Place of Origin: Britain
MOQ: 1
Payment Terms: L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability: 311/pcs/pre
Delivery Time: 3-5work days
Packaging Details: In tray
Temperature range: -30 ℃ to 50 ℃
Pressure range: Kpa 80~120
Humidity range: % RH 15 to 90
Storage life: 3-20℃ sealed storage life (months) 6
Load resistance: ω to 100
Weight: g & LT; 13
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

H2S-B4 High Resolution Hydrogen Sulfide Sensor (H2S sensor) Is Used To Monitor Trace Hydrogen Sulfide In Air Quality

Product Description:

H2S-B4 High Resolution Hydrogen Sulfide Sensor (H2S sensor) Is Used To Monitor Trace Hydrogen Sulfide In Air Quality


Features:

1. Hydrogen sulfide sensor (H2S sensor)H2S-B4 Product description:

H2s-b4 is a high-resolution H2S sensor, 7 series size, capable of detecting 1ppb of H2S gas, ideal for use in ambient air quality monitoring systems and instruments.

Second, hydrogen sulfide sensor (H2S sensor)H2S-B4 typical applications

Air quality trace hydrogen sulfide monitoring, hydrogen sulfide gas transmitter

Three, hydrogen sulfide sensor (H2S sensor)H2S-B4 main parameters

Sensitivity 2ppmH2S, nA/ PPM 1450~2050

Reaction time T90 from zero to 2ppmH2S (s) & LT; 45

Zero current 20℃ in zero air output -250~0

Interference average interference (PPB) 1

Detection value PPB 1

Range H2S quality assurance detection range (PPM) 100

Linearity full scale error, 0-2ppm linear & LT; Plus or minus 4

Overload to gas pulse stable response PPM 200

Four, hydrogen sulfide sensor (H2S sensor)H2S-B4 life

The annual variation of PPM & LT in zero drift laboratory air; + 100

Sensitivity drift percentage of annual change in laboratory air < 20

Working life output down to 80% of the original signal time (months) (warranty 12 months) 24


Specifications:

Sensitivity50ppmNH3, nA/ PPM 25~45
Reaction timet90 from zero to 50ppmNH3(s) < 60
Zero currentPPM content & LT in zero air; 7
Resolution average noise (PPM)& LT; 0.3