Categories Oil Analysis Equipment
Brand Name: YUYANG
Model Number: 4530F
Certification: ISO
Place of Origin: CHINA
MOQ: 1 Set
Price: USD13000
Supply Ability: 100 Sets per month
Delivery Time: 15 days
Packaging Details: Plywood case
Keyword: Atomic Absorption Spectrophotometer
Wavelength Range: 190 ~ 900nm
Wavelength accuracy: ±0.15nm
Warranty: 1 Year
Size: 700mm×550mm×450mm
Lamp turrets: 8 PCS
Spectrum bandwidth: 0.1nm, 0.2nm, 0.4nm, 1.0nm and 2.0nm
Power supply: 220V±22V AC
Weight: 75kg
Operation: PC controlled
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
  • Submit Buying Request
    • Product Details
    • Company Profile

    4530F PC Controlled Atomic Absorption Spectrophotometer

    4530F PC Controlled Flame Graphite Furnace Atomic Absorption Spectrophotometer

    Features:
    This AAS machine is complete automatic controlling system;
    The AA spectrophotometer can be worked with advanced graphite furnace temperature controlling technology;
    Safe, reliable and convenient flame system;
    Multi-functional software workstation;
    This flame system atomic absorption spectrophotometer is ompletely controlled by PC, can flexibly select the flame and graphite furnace atomizer (optional);
    Integrated floated optical platform design can obviously improve the optical system shock resistance and can keep stable though use the optical signal for a long term;
    Eight light stands can be changed automatically and preheat the eight element lights meantime as well as optimize the working condition of the hollow cathode lamp;
    Position adjusting: the best height of the flame burner and can automatically set the front and rear positions;
    Fully automated wavelength scanning and peak searching;
    Complete safety chains protection equipment: the function of warning and automatic safety protection towards the wrong burner, leakage of the gas, under voltage of air and the abnormal flameout;
    Deuterium lamp and self-absorption background regulation;
    Data processing: super strong database, possesses more than 500 data self-storage and cut-off storage function, can store the analyzed result with the form of EXCEL and the testing method and the result can be randomly called;
    Measuring method of AAS machine: flame absorption method and emission method;
    Result printing: parameter printing, data result printing and diagram printing;


    Instrument Configuration:

    Atomic Absorption Spectrophotometer 1pc
    Air compressor without noise 1pc
    PC workstation: computer, HP inkjet, apparatus operating software (the object can be prevailed on the specific configuration)
    A set of accessories (including Acetylene reducing valve, one copper and manganese element lamp and air filter)


    Specifications:

    Wavelength range190 ~ 900nm
    Spectral bandwidthCan automatically change between 0.1nm, 0.2nm, 0.4nm, 1.0nm and 2.0nm
    Accuracy of the wavelength±0.15nm
    Receptivity of the wavelength≤0.04nm
    Base line stability≤0.002A/30min (Cu)
    Characteristic viscosity0.02μg/ml/1% (Cu)
    Checking limit0.004μg/ml (Cu)
    Preciseness0.5%
    Grating1800 lines/mm
    InflamerAll-metal titanium burner
    AtomizerEffective glass atomizer
    Lamp standEight lamps
    D2 background correction abilityWhen the background is 1 A, the background ability should be deducted not less than 50 times; self absorption background deduction method
    Size and weight700mm×550mm×450mm 75Kg
    Power source220V±22V AC


    Flame System:

    Acetylene air burner100mm
    Nitrous oxide acetylene flame head50mm
    Ignition dynamic baseline drift≤0.006A/30min
    (Cu) Characteristic viscosity≤0.025μg/ml/1%
    Related standard deviation of the accuracy≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)
    Safety systemCan automatically cut off the gas when the pressure is not enough, the power is off, flameout and unconformity of the burner


    Graphite Furnace:

    The highest temperature3000℃
    The largest temperature rising speed≥2000℃/S
    Characteristic quantityCd≤0.5×10-12g Cu≤0.5×10-11g
    AccuracyCu≤3% Cd≤3%
    Size and weight550mm×450mm×300mm 65Kg
    Safety systemOver current protection
    ProtectionLow air pressure alarm, low cooling water flow alarm
    Power source and power220V±22V AC 7000W